Campos,T.V.O.; Jacobovitz,S.; Almeida,H.G.; Massensini,A.R.; Moraes,M.F.D.. |
Several methods have been described to measure intraocular pressure (IOP) in clinical and research situations. However, the measurement of time varying IOP with high accuracy, mainly in situations that alter corneal properties, has not been reported until now. The present report describes a computerized system capable of recording the transitory variability of IOP, which is sufficiently sensitive to reliably measure ocular pulse peak-to-peak values. We also describe its characteristics and discuss its applicability to research and clinical studies. The device consists of a pressure transducer, a signal conditioning unit and an analog-to-digital converter coupled to a video acquisition board. A modified Cairns trabeculectomy was performed in 9 Oryctolagus... |
Tipo: Info:eu-repo/semantics/article |
Palavras-chave: Glaucoma; Intraocular pressure measurement; Trabeculectomy; Signal processing; Computer-assisted measurement. |
Ano: 2006 |