Registro completo |
Provedor de dados: |
BJPS
|
País: |
Brazil
|
Título: |
Comparative performance of lab tests and blood testing device to monitor glucose, total cholesterol and triacylglycerol in type 2 diabetic patients
|
Autores: |
Zubioli,Arnaldo
Batista,Marcia Regina
Stefanello,Talitha Fernandes
Carrara,Marcia Aparecida
Bazotte,Roberto Barbosa
|
Data: |
2011-09-01
|
Ano: |
2011
|
Palavras-chave: |
Laboratory tests
Glycated hemoglobin
Glycemia
Total cholesterol
Triacylglycerol
Type 2 diabetes
|
Resumo: |
The performance of lab tests (LT) and blood testing devices (BTD) to monitor glycemia vs. glycated hemoglobin A1c (A1c) were compared. In addition, the performance of blood glucose, total cholesterol (TC) and triacylglycerol measured by LT and BDT were compared. All parameters were measured based on the same blood samples from overnight fasted type 2 diabetic patients (T2DP). Linear regression analysis was used for all comparisons. The results showed that A1c correlated better with LT-glucose (r = 0.58) than BTD-glucose (r = 0.42). Moreover, LT vs. BTD showed r values of 0.90, 0.82 and 0.92 for glucose, TC and triacylglycerol, respectively. It was concluded that the performance of LT-glucose was better than BDT-glucose. Moreover, since triacylycerol and TC measured by BTD correlated better with LT compared to BDT-glucose vs. LT-glucose, the inclusion of BTD-TC and BTD-triacylglycerol for detecting and monitoring hyperlipidemia in T2DP should be considered.
|
Tipo: |
Info:eu-repo/semantics/article
|
Idioma: |
Inglês
|
Identificador: |
http://www.scielo.br/scielo.php?script=sci_arttext&pid=S1984-82502011000300020
|
Editor: |
Universidade de São Paulo, Faculdade de Ciências Farmacêuticas
|
Relação: |
10.1590/S1984-82502011000300020
|
Formato: |
text/html
|
Fonte: |
Brazilian Journal of Pharmaceutical Sciences v.47 n.3 2011
|
Direitos: |
info:eu-repo/semantics/openAccess
|